Cleaning And Maintenance; Cleaning Device Surfaces; Maintenance Schedule; Restarting The System - Sartorius Flexsafe Pro Operating Instructions Manual

Mixer drive unit
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 3

Cleaning and Maintenance

48
Operating Instructions Flexsafe
9
Cleaning and Maintenance
9.1

Cleaning Device Surfaces

Requirements
The device is removed from the Palletank
Ending the Mixing Process", page 45).
Procedure
CAUTION Risk of injury due to uncontrolled rolling away of the
t
device! Place it on a level surface and secure the brakes. Wear personal
protective equipment.
Turn the device off.
t
Disconnect the power supply cable from the power supply at the
t
installation site.
If the device is connected to an external control system or other system:
t
Remove the connection cable and close the connection with the closing
cap.
NOTICE Corrosion or damage to the device due to unsuitable cleaning
t
agents!
Do not use corrosive, chloride-containing, and aggressive cleaning
t
agents.
Do not use cleaning agents that contain abrasive ingredients, e.g.
t
scouring agents, steel wool.
Do not use solvent-based cleaning agents.
t
Only use suitable cleaning agents (materials see Chapter "15.8
t
Materials", page 59) and observe the product information for the
cleaning agent used.
WARNING Risk of injury due to hot surfaces! Allow the motor to cool
t
down.
Wipe the housing with a slightly damp cloth. Use a mild soapy solution or
t
a suitable cleaning agent for more severe contamination.
9.2

Maintenance Schedule

Interval
Component
12 months
Device
12 months
Sensors
9.3

Restarting the System

Procedure
Commission the device (see Chapter "6 Getting Started", page 30).
t
Pro Mixer Drive Unit
®
for Mixing (see Chapter "8.12
®
Action
Have the function of the device
checked by Sartorius Service.
Have the device sensors cali-
brated by Sartorius Service.

Hide quick links: