11–3 Error/Warning Code Details
 Error/warning codes cannot be used depending on the version of the controller.
<Error Code Details>
Code
Contents
E1
Excess Current
E2
Over Voltage
E3
Motor Sensor Malfunction
E4
CONTROLLER Overheat
E5
Break Circuit Trouble
E6
Rotor Lock
E7
Low Air Pressure
E8
Over Load
Communication
E9
Interception
External Control Signal
EA
Error
EC
Internal Memory Error
EE
Emergency Stop Error
FAN Malfunction
EF1
(80 Square Type)
FAN Malfunction
EF2
(40 Square Type)
EH
Over Speed
EL
Incompatible Motor
Motor Power Line
EP
Disconnected
Parameter " P8 " Setting
EFP
Error
Et
Motor Overheat
Unknown Error
Error Occurred
<Warning Code Details>
Code
Contents
Motor Cord
A0
Low Air Pressure
A1
CONTROLLER Overheat
A2
Over Load
A3
Emergency Stop Signal
A4
Over Air Pressure
A5
Motor Overheat
A6
Motor Power Line
A7
AF Fan Stopped
AF
Spindle Operation during
FAN Malfunction
Unknown Warning
Warning Occurred
NOTICE
Motor Current beyond safe limits.
Motor Voltage beyond safe limits.
The sensor signal has malfunctioned or Motor Cord Connector is not
connected.
CONTROLLER overheat.
Trouble with the Brake Circuit.
Motor Stalled for more than 3 seconds.
・Inadequate air pressure is supplied at motor startup.
・This inadequate air pressure status continues for four seconds or more
during motor rotation.
Torque limits are exceeded for too long a period of time.
Intercept communication with SELECTOR.
(Only if using CONTROLLER connect to E3000 SELECTOR.)
・When Control Mode is in AUTO, the Control Command Signal is 'ON
(Closed)' before Main Power Switch ⑨ is turned ON.
・When Control Mode is AUTO, the ERROR command is released without
stopping the Control Command Signal 'OFF (Open)'.
Internal Memory Problem (EEPROM).
・Activated when the Emergency Stop Signal is OFF (Open).
・During rotation, an emergency stop occurred by the Emergency Stop
Signal turning OFF (Open).
FAN has Stopped (80 Square Type).
FAN has Stopped (40 Square Type).
Rotating Speed is beyond the motors capability.
An unrecognizable motor is connected to the CONTROLLER.
Motor Cord (Power Line) Connector is not connected.
A mistake has been made while setting Parameter " E8 ".
The motors Internal Temperature has risen above an acceptable amount.
An unknown error has occurred.
An error has occurred.
Motor Cord or Connector are not connected or damaged.
Low Air Pressure during motor rotation.
CONTROLLER Overheat.
Motor Torque Load exceeding safe limits.
Emergency Stop Signal 'OFF (Open) ' in Emergency Stop Mode Condition.
Excessive Air Pressure.
The inside temperature of the motor has reached the warning level.
Motor Power Line or Connector not connected or damaged.
The fan has stopped but the motor is temporarily operational.
While the warning code "AF" is displayed, the buzzer will sound.
An unknown warning has occurred.
A warning has occurred.