Siemens KION Service Manual page 91

Anesthesia workstation
Table of Contents

Advertisement

KION / KION-i
MGA-SW
The MGA-SW for the optional multigas analyzer is
stored in the multigas analyzer. There are two
different multigas analyzers that can be used:
• KION MultiGas 2000
• KION MultiGas+.
There are different softwares as well as software
installation procedures for the two versions.
KION MultiGas 2000
The KION MultiGas 2000 software is stored in two
FLASH memories and can be downloaded via a PC
connected to the analyzer. The two softwares in the
KION MultiGas 2000 are:
• PrC-SW (Protocol Converter Software)
• AION-S (Gas Analyzer Software).
KION MultiGas+
The MultiGas+ software is stored in two
exchangable program memories on the MGM S
B
inside the analyzer. The two softwares in the
OARD
KION MultiGas+ are:
• AMA-PROM (Agent Measurement Analyzer)
• AIDA-PROM (Agent Identification Analyzer).
For further technical information regarding the
multigas analyzers, refer to the Service Manual for
the concerned unit (KION MultiGas 2000 or KION
MultiGas+).
E382 E392E 061 01 02 02
Connecting a PC to the KION/KION-i
A PC can be connected to the KION / KION-i
workstation, e.g. to download diagnostic logs.
In the description below, a PC with the software
HyperTerminal (included in Windows) is used.
Connection and setup
Cable connection
• Make sure that the PC and the KION / KION-i are
turned Off.
• Connect an RS232 cable (Order No. 47 14 346
E530U) to a COM port on the PC.
• Connect the other end of the cable to the KION /
KION-i as follows:
– For access to the "Diagnostic Log", connect to
the X3a port.
– For access to the "CPS/IDS Diagnostic Log",
connect to the X4 port.
• Turn the PC On and the KION / KION-i to Standby.
First time-setup
YSTEM
If this is the first time the PC is used for
communication with the KION / KION-i, terminal
settings must be made. These settings can be saved
and used whenever the PC is connected to a KION /
KION-i.
• Start HyperTerminal on the PC.
• Select a name (e. g. KION) and an icon. Click OK.
• Select the used COM port in the "Connect to"
dialog box. Click OK.
• Make the following COM port-settings:
Bits per second:
Data bits:
Parity:
Stop bits:
Flow control:
• Click OK.
• The HyperTerminal window will now be open and
active.
• Select File/Save in the HyperTerminal menu to save
the settings and thereby create a HyperTerminal
setting to be used whenever the PC is connected
to a KION / KION-i.
• Press the Enter key. The "System Console Main
Menu" will now appear in the HyperTerminal
window (if a sub-menu is displayed, return to the
"System Console Main Menu"). Communication
with the KION / KION-i is established.
Siemens-Elema AB
Service procedures
19.200
8
None
1
None
5
5 - 19

Advertisement