Disassembling and assembling
#
4
4 - 26
A
B
"
!
#
• To remove the safety valve SV5–SV7 (13), loosen
the nut (14) using a 19 mm open-end spanner and
unscrew the safety valve.
• When mounting the safety valve:
!
– Screw the nut (14 in Fig. B) as much as possible
onto the safety valve. Note: There is a groove for
the gasket in the nut. This groove must be facing

upwards as showed in Fig. B.
– Place a new gasket (15 in Fig. C) in this groove.
– Screw the valve (13) as much as possible into the
manifold.
– Unscrew the valve until the two outlet holes (16)
are facing away from the operator (downwards/
inwards).
– Hold the valve in this position and secure it by
tightening the nut (14). Tightening torque for the
C
nut is approx. 35 Nm.
Siemens-Elema AB
KION / KION-i
E382 E392E 061 01 02 02