Troubleshooting
MultiGas O2 Failure (562)
Technical O2 module Alarm from the CAN/RS232
interface followed by the clarifying diagnostic
message:
• 605 – MGI: O2_MODULE_ERROR
KION ALARM DISP ERR (830)
Technical Display Alarm from ALA followed by one of
the clarifying diagnostic message:
• 032 – ALA: DIAG_ALARM_BUZZER_ERROR
• 033 – ALA: DIAG_STARTUP_LED_TEST_FAILED
• 049 – ALA: DIAG_ERROR_LINE_ERROR
MGI: T_COMM_TIMEOUT (960)
6
Technical Communication Alarm from the CAN/
RS232 interface followed by one of the clarifying
diagnostic messages:
• 606 – MGI: DIAG_COMMAND_TIMEOUT_ERROR
• 608 – MGI: DIAG_STATUS_TIMEOUT_ERROR
6 - 24
Multigas analyzer logs
Multigas analyzer serial number
Each time the power switch is turned from Off to
Standby or On, a Diagnostic Log including serial
number of the built-in multigas analyzer is created. If
the multigas analyzer serial number written in the log
vary between different start-up occations, this
indicates that the multigas analyzer has been
replaced. However, it may also indicate that the
KION Monitor has been moved between different
KION workstations and thus been connected to
different multigas analyzers.
Multigas analyzer information in the logs
Some logs without text in the Description-field can
be identified as multigas analyzer error logs by the
information written in the Diagnostic number- / Event
ID-field.
The table below shows the Diagnostic numbers /
Event IDs related to multigas analyzer errors.
8038b016
MGM Status: Persistent Pressure Data Invalid
8038b01d
MGM Status: Self Test Error
8038b01f
MGM Status: IR Signal Degraded
8038b020
MGM Status: 4710 Operation Fault
8038b021
MGM Status: 4710 Data Accuracy Not Guaranteed
8038b022
MGM Status: 4710 Data Not Available
8038b023
MGM Status: Check 4710 Head
8038b024
MGM Status: 4750 System Fault
8038b025
MGM Status: 4750 Agent ID Module Operation
Fault
8038b026
MGM Status: 4750 Agent ID Module Accuracy
Not Guaranteed
8038b027
MGM Status: 4750 Agent ID Module Data Not
Available
8038b029
MGM Status:Power Error
8038b01e
4710 Operation Fault + Set Synch Error
80367001
CAN_HW_Overun
8038b006
Cannot select MGM agent 12857.
Action: Upgrade to VF2.3 or later.
For qualified troubleshooting of the multigas
analyzers, always use the recommended service
software tools (available on the Intranet).
In case of multigas analyzer errors, always check that
KION Monitor-SW and CPS/IDS-SW according to
applicable Release Notes are installed.