Siemens KION Service Manual page 160

Anesthesia workstation
Table of Contents

Advertisement

x
312
... 01 38 ...
x
313
... 01 39 ...
x
314
x
320
... 01 40 ...
100
x
323
... 01 43 ...
100
x
329
... 01 49 ...
100
PAN: DIAG_CAN_SEND_TIMEOUT
PAN:
DIAG_CAN_WARNING_LIMIT_REACHED
PAN:
DIAG_CAN_WARNING_LIMIT_OK_AGAIN
PAN:
DIAG_SETTING_VOLUME_ALARM_LIMIT_ER
ROR
PAN:
DIAG_SETTING_HIGH_PRESSURE_LIMIT_E
RROR
PAN:
DIAG_SETTING_BAROMETER_PRESSURE_
ERROR
Fault: Communication bus (CAN) is not working properly.
Description: The CAN driver reports a send timeout error.
What to do:
1.
Replace PC 1672 Control Panel.
2.
If this log is repeted several times; check CAN communicating units, i.e.
microprocessor controlled PC boards, multigas analyzer, KION Monitor and
DuoView. Also check PC board- and cable connectors.
3.
Report to EM-HSC.
Fault: Communication bus (CAN) is not working properly.
Description: At least one error counter has reached warning limit. Refer to diagnostic
log 314.
What to do:
1.
If this log is repeted several times; check CAN communicating units, i.e.
microprocessor controlled PC boards, multigas analyzer, KION Monitor and
DuoView. Also check PC board- and cable connectors.
2.
Report to EM-HSC.
Information: Communication bus (CAN) is working again.
Description: At least one error counter has reached warning limit. The problem is
solved by the HW. Refer to diagnostic log 313.
Fault: The Lower Vol. Alarm Limit potentiometer is not working properly.
Description: The reading of the potentiometer failed.
What to do:
1.
Check / replace potentiometer.
2.
Replace PC 1672 Control Panel.
Fault: The Upper Pressure Limit potentiometer is not working properly.
Description: The reading of the potentiometer failed.
What to do:
1.
Check / replace potentiometer
2.
Replace PC 1672 Control Panel.
Fault: The Barometer is not working properly.
Description: The reading of the barometer failed or the read value is out of range
(600 < Barometer Pressure < 1150).
What to do: Replace PC 1672 Control Panel.