Troubleshooting
Docking station
silicone muff
6
6 - 12
Test case 6: Patient cassette docking station
1. Remove the docking station silicone muff
including the 3 bacteria filters.
2. Clean the silicone muff if required.
3. Check / replace the bacteria filters.
4. Reinsert the silicone muff gently to prevent
damages to the bacteria filters.
Test case 7: Pressure transducers including
tubings
1. Check for obstructions or poor connection in the
pressure transducer tubes. Check tube
connections on the pressure transducers and in
the docking station.
2. Replace complete PC 1750 Transducer (old
version) or the concerned PC 1781 Pressure
transducer (new version).