8. PC 1733 / PC 1796 Interconnection
The PC 1733 / PC 1796 I
NTERCONNECTION
connects a number of units in the KION workstation.
The only electronic functions on this PC board are
output filters and PTC resistors.
There are two versions of this PC board:
• PC 1733 on units with S/N 00125–01405.
• PC 1796 on units with S/N 01501 and higher.
Four PC boards are mounted in PC board connectors
on the upper side of the PC 1796:
• PC 1730 C
ONTROL
• PC 1737 M
EASURING
• PC 1750 T
RANSDUCER
• PC 1797 F
G
F
RESH
AS
LOW
With PC 1733 I
NTERCONNECTION
Instead, PC 1748 F
G
RESH
connected to PC 1733 via a cable.
The under side of PC 1733/PC 1796, accessible
behind a trolley cover, contains cable connectors for:
• Patient unit cable.
• Pneumatic section cable.
• Control panel cable.
• Pressure transducers on the optional B
S
M
.
UPPLY
ANIFOLD
The PC board inputs/outputs are described in chapter
"Diagrams".
E382 E392E 061 01 02 02
electrically
.
, PC 1797 is not used.
F
is used and
AS
LOW
G
ACKUP
AS
Siemens-Elema AB
9. PC 1730 Control
Controls flow, pressure, composition and concentra-
tion in the gas delivered to the patient. All ventilation
mode functionality is controlled by this unit.
PC 1730 C
contains a microprocessor that
ONTROL
executes the CON-software (CON-SW) that is stored
in a FLASH-PROM. New versions of the CON-SW
can be installed either by replacing PC 1730 or by
using the Software download tool.
The main functions of PC 1730 C
• General parameters (used or created by CON).
• Fresh gas production (excluding control of
anesthetic agent type and concentration).
• Inspiratory fresh gas control.
• Drive gas for rebreathing.
• Regulation of expiration.
General parameters
Some of the general parameters used or created by
CON are:
• Pressure: Input pressure signals from the four
pressure transducers on PC 1750 are supplied via
the A/D C
for regulation of inspiration and
ONVERTER
expiration.
• Timing: Output Insp. time and Exp. time signals
are generated. These are used by MEA to control
zeroing of the flow transducers.
• Fresh gas outlet: Input signal from the Hall sensor
in the A
F
G
UXILIARY
RESH
AS
valve is opened. This signal is used to light up a
LED on the C
P
ONTROL
ANEL
• Pressure alarm: Input signal from MEA indicating
that upper pressure limit is reached. Used to stop
inspiration time and immediately start expiration
time. This will open the PEEP V
other valves in safe positions.
Description of functions
are:
ONTROL
O
indicates when the
UTLET
.
and to set the
ALVE
3 - 11