KION / KION-i
E382 E392E 061 01 02 02
Vaporizer magazine
Replace:
P. O-ring 9.3 x 2.4 for the vaporizer magazine.
Apply a thin layer of special grease on the O-ring.
P
Q
Q. O-ring 36.2 x 3.0 for the vaporizer magazine.
Apply a thin layer of special grease on the O-ring.
R. One-way valves for the vaporizer magazine inlet
connectors.
R
S. O-rings 14.1 x 1.6 for the vaporizer magazine inlet
connectors.
S
T. O-rings 14 x 3 for the vaporizer magazine inlet
and outlet connectors.
T
U. Screw covers for the vaporizer magazine. Not
shown in the illustration.
PIVap vaporizer
Replace:
V. O-rings 5.3 x 2.4 for the vaporizer valve. Apply a
V
thin layer of special grease on the O-rings.
Replace this O-ring on each one of the vaporizers
to be used on the concerned workstation. Three
O-rings are included in the kit, order further O-
rings if required.