Description of functions
25. Backup gas supply manifold
With the optional KION B
backup gas cylinders can be connected to the KION /
KION-i workstation. The manifold is intended for:
• N
O (one gas cylinder)
2
• AIR (one gas cylinder)
• O
(two gas cylinders)
2
The B
G
S
ACKUP
AS
UPPLY
PIN-index connections (yokes) for 4 gas cylinders:
• N
O – One 3-liter gas cylinder. Max cylinder gas
2
pressure is 70 kP x 100 (70 bar) at 35°C (95°F)
normally correponding to approx. 50 kPa x 100
3
(50 bar) at 20°C (68°F) before the pressure
regulator.
• AIR – One 3-liter gas cylinder. Max cylinder gas
pressure is 200 kP x 100 (200 bar) before the
pressure regulator.
• O
– Two 3-liter gas cylinders. Max cylinder gas
2
pressure is 200 kP x 100 (200 bar) before the
pressure regulator.
The gas inlet channels, inside the yokes, contains
inlet filters and one-way valves (OV5 – OV8). The
filters must be replaced during the "Twelve months
maintenance".
The cylinder gas pressure is reduced by the preset
pressure regulators (REG5 – REG7). The gas is then
connected to the G
AS
tubing.
The integrated pressure transducers are connected
to PC 1733 / PC 1796 I
pressure in the cylinders will be displayed on the
KION Monitor. Alarm is activated if the gas pressure
is outside the preset alarm limits.
There are three safety valves (SV5 – SV7) on the
manifold; one for each gas. The safety valves are
preset to open at a pressure of 6.5 kPa X 100
(6.5 bar).
For further information, refer to the "KION Optional
Accessories Operating Manuals".
3 - 26
G
S
M
ACKUP
AS
UPPLY
ANIFOLD
M
is equipped with
ANIFOLD
D
B
via internal
ISTRIBUTION
LOCK
, thus the gas
NTERCONNECTION
Siemens-Elema AB
26. Suction equipment
,
The optional KION S
on the K
System.
ION
The main parts of the KION S
• S
E
that creates the vacuum. The
UCTION
JECTOR
S
E
is powered with drive gas (AIR)
UCTION
JECTOR
from the G
D
AS
ISTRIBUTION
OFF switch and a vacuum control knob on the
S
E
.
UCTION
JECTOR
• S
B
. The Siemens 1.8 liter bottle is
UCTION
OTTLE
recommended.
For further information, refer to the "KION Optional
Accessories Operating Manuals".
27. Auxiliary O
The optional KION A
Oxygen Flowmeter) can be mounted on the KION /
KION-i workstation. The flowmeter is supplied with
O
from the G
D
AS
ISTRIBUTION
2
The flow through the A
the flowmeter knob and the actual flow is indicated
by the measuring ball inside the measuring channel.
The measuring range of the flowmeter is 0–15 l/min.
For further information, refer to the "KION Optional
Accessories Operating Manuals".
28. Gas evacuation
The optional G
E
AS
VACUATION
the KION / KION-i workstation is the KION evac
system.
The KION evac system is a Servo Evac 180 equipped
with a bracket and a holder arm to be mounted onto
the KION / KION-i workstation.
The KION evac system is connected to the EVAC
outlet on the KION / KION-i workstation. A Siemens
evacuation ejector or the hospital's central vacuum
system may be used as an active suction source.
For further information regarding function, operation
and technical specifications, see the "Servo Evac 180
– Operating Manual".
KION / KION-i
E
can be mounted
UCTION
QUIPMENT
E
UCTION
QUIPMENT
B
. There is an ON/
LOCK
supply
2
O
S
(Auxiliary
UXILIARY
UPPLY
2
B
.
LOCK
O
S
is set with
UXILIARY
UPPLY
2
equipment to be used on
E382 E392E 061 01 02 02
are: