Internal Communication - Siemens KION Service Manual

Anesthesia workstation
Table of Contents

Advertisement

KION / KION-i

Internal communication

The diagram below shows the basic electrical signals
between the different electronic subsystems:
PAN
MEA
CON
= CAN node
CAN bus
Internal communication via CAN bus protocol (CAN =
Controller Area Network).
The CAN bus is a simple two-wire differential serial
bus system. The CAN bus operates in noisy electrical
environments with a high level of data integrity.
TAXI bus
Communication between the CPS/IDS and the KION
Monitor.
E382 E392E 061 01 02 02
CAN bus
Emergency line
ALA
Standby
Reset
TAXI bus
MON
MGI
Reset
PWR
Emergency line
Digital signal in separate cabel for alarm distribution
in case of CAN bus malfunction.
Standby
Digital signal in separate cabel that activates standby
functions.
Reset
Digital signal in separate cabel that ensures a
controlled start of the microprocessors at a stable
voltage level.
Siemens-Elema AB
Introduction
KION
Monitor
CPS / IDS Diagnostic
SC 9000 Diagnostic
VGA output
CIS (RS232)
Recorder
Infinity Network
MIB 1 & 2
RS232 Interface
(multigas analyzer)
Mains power
Power outputs (Optional
isolation transformer)