MGI:
DIAG_PATIENT_DATA_TIMEOUT_ERROR
MGI: DIAG_TIMER_ERROR
MGI: DIAG_WD_ERROR
MGI: DIAG_RESTART_ERROR
Fault: The multigas analyzer has stopped sending patient data.
Description: No patient data is received from the multigas analyzer within 1.5
seconds. The multigas analyzer should send patient data with 1 Hz.
What to do:
1.
Communication error between the multigas analyzer and PC 1747. Check
RS232 cable and connectors.
2.
Troubleshoot the multigas analyzer.
Fault: Hardware failure.
Description: Timer is not working properly during startup hardware test. Alarm 560 is
generated.
What to do: Replace PC 1747 CAN/RS232 Interface.
Fault: Hardware failure.
Description: Watchdog is not working properly during startup hardware test. Alarm
560 is generated.
What to do: Replace PC 1747 CAN/RS232 Interface.
Fault: Hardware failure.
Description: The COLD_START_FLAG is not set during startup hardware test.
What to do: Replace PC 1747 CAN/RS232 Interface.