Siemens KION Service Manual page 21

Anesthesia workstation
Table of Contents

Advertisement

KION / KION-i
Batteries (spare parts) are not supplied by Siemens.
They must be purchased locally. Tested batteries for
use in the KION / KION-i workstation are listed below.
No other batteries must be used.
• CSB EVX-12260
(http://www.csb-battery.com)
• Newmax FNC 12240
(http://www.newmaxbattery.co.kr)
• Power-Sonic PS-12260 NB
(http://www.power-sonic.com)
• Hitachi HP24-12
For further information regarding the batteries and
battery suppliers, refer to the Internet address stated
at each battery type above or to the Siemens Medical
Solutions, Customer Service Intranet-Portal at
http://cs.med.siemens.de.
3. Extra internal battery (optional)
E
I
B
(two serial connected 12 V
XTRA
NTERNAL
ATTERY
sealed lead acid batteries) for extended backup time.
Gives, together with the I
backup time of approx. 120 minutes with full charged
battery.
For information regarding battery type, function and
service interval, see section 2. "Internal battery"
above.
There are two versions of the battery cable; one
version to be used if a standard I
connected and one version to be used if an E
I
B
is connected. Thus, if an extra E
NTERNAL
ATTERY
I
B
is installed, the battery cable must
NTERNAL
ATTERY
be replaced.
4. Power & Communication Interface
The following functions are included in the P
C
I
:
OMMUNICATION
NTERFACE
• Communication ports – PC 1754 C
PC 1856 C
.
ONNECTOR
• PC 1761 MIB C
ONNECTOR
C
2.
ONNECTOR
• C
I
, CPS network PCB alt.
OMMUNICATION
NTERFACE
IDS network PCB.
• PC 1747 CAN/RS232 I
NTERFACE
• PC 1753 DC/DC C
ONVERTER
V and a fused +24 V for power supply of the
multigas analyzer.
• PC 1675 P
S
. Voltage supply and battery
OWER
UPPLY
charge/capacity check. Generates +15 V, - 15 V,
+5 V and 24 V for the KION system.
• B
D
F
.
ATTERY
RAWER
AN
E382 E392E 061 01 02 02
B
, a total
NTERNAL
ATTERY
B
is
NTERNAL
ATTERY
XTRA
XTRA
&
OWER
alt.
ONNECTOR
alt. PC 1843 MIB
.
. Generates +12 V, +13
Siemens-Elema AB
PC 1754 Connector / PC 1856 Connector
Communication ports for connecting the KION /
KION-i workstation to network, service options, etc.
The communication ports are integrated parts of the
PC board.
There are two versions of the PC board:
• PC 1754 C
for KION with CPS.
ONNECTOR
• PC 1856 C
for KION / KION-i with IDS.
ONNECTOR
Pin configuration and signal names can be found in
chapter "Diagrams".
PC 1761 MIB Connector / PC 1843 MIB Connector 2
Converts monitoring parameters to MIB protocol.
MIB = Medical Information Bus.
There are two versions of the PC board:
• PC 1761 MIB C
ONNECTOR
• PC 1843 MIB C
ONNECTOR
IDS.
Pin configuration and signal names can be found in
chapter "Diagrams".
Communication interface
There are two versions of the C
board:
• For KION with S/N 03645 or lower: PC A101CPS
N
PCB including the small PC A301 M
ETWORK
O
S
M
PTION AND
ETUP
EMORY
are also included in the Siemens patient monitor
Communication/Power Supply module.
• For KION / KION-i with S/N 03646 or higher:
PC A110 IDS N
PCB and PC A120 CAN/MIB.
ETWORK
This PC board is also included in the Siemens
Infinity Docking Station module.
The CPS/IDS software (CPS-SW alt. IDS-SW) is
installed from a PC Card via the KION Monitor.
For further information regarding the Communication/
Power Supply module or Infinity Docking Station
module, refer to e. g. SC 7000/9000XL or SC 9000
Service Manuals.
PC 1747 CAN/RS232 Interface
Protocol converter between KION (CAN) and the
multigas analyzer (RS232).
The MGI-software is stored in an exchangable
program memory (MGI-PROM) mounted on PC 1747
CAN/RS232 I
.
NTERFACE
Description of functions
for KION with CPS.
2 for KION / KION-i with
OMMUNICATION INTERFACE
ONITOR
B
. These PC boards
OARD
3 - 3
3