4.1.2 General Installation Instructions

Hazardous voltage may be present.
Always isolate the electrical power supply before making or changing connections
to the unit. To remove the power it is recommended to use an isolating device.
Hazardous voltage is present until 60 sec after switch off. Do not handle the
controller during this time.
The controller housing must be properly grounded. Use the specified screws on
the feet of the controller housing.
Incorrect assembling of the "POWER INPUT" connector
short circuit possible.
Assure that the pin assignments of the "POWER INPUT" connector are correct
before it is plugged in.
1.
Connect the protective earth wire with a crimp-type end on the specified earth screw (see
Figure 28) on the feet of the controller (see also protective earth labels on controller).
2.
Connect the two motor connectors (sensor and power) to the controller. Assure that the
"POWER OUTPUT" connector from the motor is correctly aligned with the connector of the
controller before it is plugged in.
3.
Connect the AC power input connector. Make sure that the pin connections are correct:
- 1 x 200-240V (1-phase)
- 3 x 200-240V (3-phase delta voltage)
- Minimum Wire Gauge = AWG 14 (cooper diameter = 1.63 mm, cross section = 2.08 mm
- For usage of external circuit breakers consult Section 3.2.5.
4.
To secure the connectors, tighten all retaining screws according to the torque specifications
in Table 11 (for LPC-2000.1) and Table 12 (for LPC-2000.2).
Figure 28: Protective earth and AC power input connection of LPC-2000 controller
PL-4021-00, Rev03, DCO# 15-235
WARNING
!
WARNING
!
L1 (
L), L2 (
N), PE (= Protective Earth)
L1 ,L2, L3, PE (lines can be switched), star-voltage = 115 – 139V AC
L
N
AC Mains
1-Phase
= 200 – 240 Vrms ± 10%
U
1-Phase
PE
L1
U
Delta
L2
AC Mains
3-Phase
L3
= 200 – 240 Vrms ± 10%
U
Delta
= 115 – 139 Vrms ± 10%
U
Star
N
PE
User Manual for BPS-2000
2
)
"POWER INPUT"
Connector
U
1-Phase
"POWER INPUT"
Connector
U
Star
www.levitronix.com
L1
L2
L3
Controller
LPC-2000
Protective Earth
PE
Screw on Housing
L1
L2
L3
Controller
LPC-2000
Protective Earth
PE
Screw on Housing