Interacoustics AD629 Instructions For Use Manual page 65

Device for diagnosing hearing loss
Table of Contents

Advertisement

Guidance and manufacturer's declaration - electromagnetic Immunity
The AD629 is intended for use in the electromagnetic environment specified below. The customer or the
user of the AD629 should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test level
Electrostatic
+8 kV contact
Discharge (ESD)
+15 kV air
IEC 61000-4-2
+2 kV for power supply
Electrical fast
lines
transient/burst
+1 kV for input/output
IEC61000-4-4
lines
Surge
+1 kV differential mode
IEC 61000-4-5
+2 kV common mode
< 5% UT (>95% dip in
UT)
for 0.5 cycle
Voltage dips, short
40% UT (60% dip in
interruptions and
UT)
voltage variations
for 5 cycles
on power supply
lines
70% UT (30% dip in
UT)
IEC 61000-4-11
for 25 cycles
<5% UT (>95% dip in
UT)
for 5 sec
Power frequency
(50/60 Hz)
3 A/m
IEC 61000-4-8
Note: UT is the A.C. mains voltage prior to application of the test level.
D-0133833-A – 2022/09
AD629 - Instructions for Use - EN
Compliance
+8 kV contact
+15 kV air
+2 kV for power
supply lines
+1 kV for input/output
lines
+1 kV differential
mode
+2 kV common mode
< 5% UT (>95% dip
in UT)
for 0.5 cycle
40% UT (60% dip in
UT)
for 5 cycles
70% UT (30% dip in
UT)
for 25 cycles
<5% UT
for 5 sec
3 A/m
Electromagnetic environment -
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
greater than 30%.
Mains power quality should be
that of a typical commercial or
residential environment.
Mains power quality should be
that of a typical commercial or
residential environment.
Mains power quality should be
that of a typical commercial or
residential environment. If the
user of the AD629 requires
continued operation during power
mains interruptions, it is
recommended that the AD629 be
powered from an uninterruptable
power supply or its battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or residential
environment.