3. Position the chassis so that there is ample space between the chassis fan intake
and exhaust vents. Blockage by walls or obstructions affects the air flow needed
for cooling.
4. Before installing a module, be sure to remove the plastic thread protectors from
the mounting screws. Before inserting the module into the chassis, back the
mounting screws out to assure that there is no interference between the screws
and the mounting rails. Inspect the chassis slot to ensure there are no bent pins on
the slot connectors.
5. Ensure the ejector handles of the module are in the unlatched (downward)
position.
6. Insert CXA-m analyzer into 4 contiguous empty slots by placing the module card
edges into the top and bottom module guides.
2. Pull up
7. Carefully Slide the module completely into the chassis. When you begin to feel
resistance, pull up on the injector/ejector handle to fully insert and latch the
module into the chassis. Refer to the figure above.
8. Tighten the screws at the top and bottom of the module front panel for both
mechanical security and to ensure proper grounding of the front panel.
Performance may suffer if the screws are not tightened properly.
9. Verify the PXI chassis fans are operable and free of dust and other contaminants
that may restrict airflow.
10. Position the chassis so that inlet and outlet vents are not obstructed.
11. Cover all empty chassis slots by using filler panels or air inlet modules. You can
find them on the Keysight Technologies website.
http://www.keysight.com/find/accessories
12. Plug in and power on the chassis.