Removing Protective Pipe; Fig. 12: Removing Transportation Protection; Fig. 13: Removing Protective Pipe - Pfeiffer Vacuum QMA 4X0 Operating Instructions Manual

Quadrupole analyzer for qmg 700 hiquad
Table of Contents

Advertisement

Fig. 12:
1 Transport protection
2 Mass filter
Procedure
1. Carefully remove the transportation protection and store it for future use.
2. Check the inside for damage and wiring short-circuits.
5.1.6 Removing protective pipe
Prerequisites
● Device is switched off
● All cables have been disconnected from the analyzer
Required tools
● Screwdriver
1
2
Fig. 13:
1 Screw (3×)
Procedure
1. Remove the 3 outer screws.
2. Pull off the protective pipe.
Removing transportation protection
3
Analyzer housing

Removing protective pipe

2
Protective pipe
1
2
3
Installation
29/52

Advertisement

Table of Contents
loading