CAUTION: Medical electrical equipment needs special precautions regarding electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the
EMC information in the XXX sections of this document. Portable and mobile RF
communications equipment can affect medical electrical equipment.
Classifications
1.
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air,
oxygen or nitrous oxide.
2.
Mode of operation: Continuous.
3.
Supply CLASS I Power Adapter see "1.5 Packing List" for power adapter detail
information.
4.
No Applied Part
5.
IPX0
6.
Pollution degree of equipment: Pollution Degree 2
7.
Overvoltage category of equipment: Overvoltage category II
8.
Operation altitude of equipment: 0-5000m
9.
Material group IIIb
10. Two MOPP insulations were provided between primary and secondary. One
MOPP insulation was provided between primary and earth.
11. Two MOPP insulations were provided between primary and secondary. One
MOPP insulation was provided between primary and earth.
12. ISOLATION ports provide one MOPP against other secondary circuits.