Diagnostics and troubleshooting

Diagnosis
Short text
no.
000
Flow Computer 1
Communication Error
001
Flow Computer 2
Communication Error
002
Cabinet door open
003
Sealing Switch
Unsealed Position
050
File Read/write Error
051
Power Supply 1 Alarm
052
Power Supply 2 Alarm
100
Flow Computer 1
Calculation Error
101
Flow Computer 1 DUT
Density Transmitter
Fail
102
Flow Computer 1 DUT
Pressure Transmitter
Fail
103
Flow Computer 1 DUT
Temperature
Transmitter Fail
104
Flow Computer 1
FLASH Fail
105
Flow Computer 1
FRAM Fail
106
Flow Computer 1 I/O
Comms Fail
107
Flow Computer 1 MM
Pressure Transmitter
Fail
108
Flow Computer 1 MM
Temperature
Transmitter Fail
109
Flow Computer 1
Invalid Hardware
Version
110
Flow Computer 1
System Restart
111
Flow Computer 1 MM
Comms Fail
48
9
Diagnostics and troubleshooting
9.1

List of error messages

Diagnostic behavior:
• Error:
• Warning:
Diagnosti
Possible cause
c
behavior
Communication between HMI and flow
computer is interrupted.
Communication between HMI and flow
computer is interrupted.
The cabinet door is open.
The sealing switch has been activated and is in
the "unsealed" position.
The HMI could not read or write to the file.
No 24 V signal available at Power Supply 1.
No 24 V signal available at Power Supply 2.
A calculation error has occurred in the flow
computer.
No signal received at analog input of DUT
density sensor.
No signal received at analog input of DUT
pressure sensor.
No signal received at analog input of DUT
temperature sensor.
An internal flow computer error has occurred.
An internal flow computer error has occurred.
An internal flow computer error has occurred.
No signal received at analog input of MM
pressure sensor.
No signal received at analog input of MM
temperature sensor.
The MM application was installed on an
incompatible flow computer.
The flow computer has been restarted.
Modbus RTU communication between flow
computer and Promass MM has failed.
Master Meter System
Remedy
Ensure that flow computer is operational and
that Ethernet cables are connected.
Ensure that flow computer is operational and
that Ethernet cables are connected.
Close cabinet door.
Set sealing switch to "sealed" position.
Restart HMI computer.
Ensure that Power Supply 1 is switched on.
Replace power supply if necessary.
Ensure that Power Supply 2 is switched on.
Replace power supply if necessary.
Restart flow computer.
If error persists, contact Endress+Hauser.
Check wiring of 4 to 20-mA signal at analog
input.
Check sensor.
Check wiring of 4 to 20-mA signal at analog
input.
Check sensor.
Check wiring of 4 to 20-mA signal at analog
input.
Check sensor.
Restart flow computer.
If error persists, contact Endress+Hauser.
Restart flow computer.
If error persists, contact Endress+Hauser.
Restart flow computer.
If error persists, contact Endress+Hauser.
Check wiring of 4 to 20-mA signal at analog
input.
Check sensor.
Check wiring of 4 to 20-mA signal at analog
input.
Check sensor.
Install application on a 3rd generation or newer
flow computer.
This is normal behavior following an intended
restart.
In any other case, contact Endress+Hauser.
Check wiring of Modbus RTU (RS-485) cables.
Ensure that Promass is switched on. Check
Modbus RTU communication settings.