Thermo Scientific 3906 Operating Manual page 73

Environmental chamber
Table of Contents

Advertisement

See
Manual
Section
--
Inspect door latch, hinges and
door gasket seal.
8
Check air exchange ventilator
caps for adjustment; open or
close as required.
9
Perform a complete
decontamination procedure.
Wipe down interior, shelves,
side panels with disinfectant.
Rinse everything well with
sterile water.
13
Verify and document all cali-
brations at minimum once a
year.
10
Inspect and clean the humidity
generator, at minimum once a
year.*
5
Clean drip pan and drain lines.
9
Clean refrigeration system con-
denser.*
8
Verify defrost cycle for below
10°C operation.
--
Change filters (under normal
conditions).
--
Replace filter elements on the
dryer, if applicable. †
--
Safety check of the electrical
system according to national
legislation.
--
Rebuild solenoid valves on the
dryer, if applicable. †
--
Replace/repack the desiccant
towers on the dryer, if applicable.
--
Replace the check valve and
o-rings on the dryer, if applicable.
--
Check in regular intervals the
light modules for defect lamps.
Thermo Scientific
Action
3 Months
Yearly
2 Years
Between Experi-
ments. More frequent
deconta- mination
may be required,
depending on use
and environmental
conditions.
Depending upon the
quality of water used
in the humidification
system, it may be
necessary to clean
more frequently.
Maintenance |
5 Years
6 Years
Model 3900 Series | 10-3