Thermo Scientific 3906 Operating Manual page 82

Environmental chamber
Table of Contents

Advertisement

| Specifications
Chapter 12
Lighting performance data per light module
for animal hatching applications
(Temperature performance with humidity control)
Evaluation surfaces on the shelf (width x depth)
Average Illumination on the shelf at 25°C
Illumination distribution relating evaluation surfaces
Electrical Data
Power line voltage (+/-10%)
Power line frequency
Power rating with all options
Measured at ambient temperature of 26°C /79°F
Max current consumption with all options
Measured at ambient temperature of 26°C /79°F
Power Cord / Plug
Earthing system (e.g. 1/N/PE)
IP protection system
Protection class
Overvoltage category according to IEC
1
60364-4-443
2
Pollution degree
Ambient conditions
Ambient Temperature Range
Ambient Temperature Range with lighting appli-
cation
Max. rel. humidity in service at or below 32°C,
non condensing
Max. rel. humidity in service at or below 26°C,
non condensing with lighting application
Storage temperature range
|
12-6
Model 3900 Series
Unit
3906
2
m
Lux
%
Unit
V
Hz
W
A
Unit
°C (°F)
°C (°F)
% r.F.
% r.F.
°C (°F)
Models
3943
259 x 218
432 x 364
610 x 510
10000
≤ ± 5 (259 x 218)
≤ ± 12 (432 x 364)
≤ ± 26 (610 x 510)
Models
3906
3943
400
50/60
3800
15.0
CEE
16 A, 6h
3/N/PE
IP 20
I
(PE- connected)
II
2
Models
3906
3943
16 - 32 (61 - 90)
18 - 26 (64 - 79)
55
70
20 - 60 (68 - 140)
400
50/60
3900
15.0
CEE
16 A, 6h
3/N/PE
Thermo Scientific

Advertisement

Table of Contents
loading

This manual is also suitable for:

Thermo scientific 3943

Table of Contents