| Product Description
Chapter 5
|
5-14
Model 3900 Series

3906 Sockets Light Module

The light module socket positions are at the right air duct wall in the chamber.
Socket position 1 (supreme socket)
rd
3
hole position of shelf holder from above
distance from the chamber top
= 6 inches ( 152 mm)
socket position below of the light module 1
Socket position 2
th
6
hole position of shelf holder from above
distance from the chamber top
= 12 inches ( 305 mm)
socket position above of the light module 2
Plug-in the mains supply and secure plug in the socket by turning clockwise.
NOTE
Sockets not in use must be covered with a locking cap to avoid moisture in the
connector!
3906 Standard positions for max. 2 light modules with equal height distances.
3943 Standard positions for max. 3 light modules with equal height distances.
One light module must be placed on top most position, otherwise the
temperature uniformity is influenced.