Warranty; Intended Use; Unintended Use - Thermo Scientific 3906 Operating Manual

Environmental chamber
Table of Contents

Advertisement

| Safety Notes
Chapter 1
|
1-4
Model 3900 Series

Warranty

Thermo Fisher Scientific warrants the operational safety and functions of the
environmental chambers only under the condition that:
The environmental chamber is operated and serviced exclusively in accordance with
its intended purpose and as described in these operating instructions.
The environmental chamber is not modified.
Only original spare parts and accessories that have been approved by Thermo
Scientific are used (third-party spares without Thermo Scientific approval void the
limited warranty).
Inspections and maintenance are performed at the specified intervals.
An operation verification test is performed after each repair activity.

Intended Use

Thermo Scientific Environmental Chambers with light option are laboratory devices
intended for
Stability testing,
Shelf life studies, packaging testing,
Light stability testing,
Cultivation of: cells, tissue, micro organism cultures for research only,
Plant growth,
Hatching of insects, fish or other animals and microorganisms that require light
exposure,
Refrigerated storage or long term storage of substances and samples.
The devices employ:
Precision temperature control - above or below and above ambient temperature,
depending on model
Precision humidity control: some models only
On option: Precision CO
On option: Light control for day/night simulation and light stability testing for
simulating the specific physiological ambient conditions for these cultures, organisms
and test or storage applications acc,. ICH guideline.

Unintended Use

The appliance is not explosion-proof. To avoid the risk of explosion do not load the
environmental chamber with tissue, material or liquids that:
are easily flammable or explosive,
control