Risk of UV exposure
Use the light module only for its intended application. Don't use ICH modules for
plant growth or animal hatching.
UV-A and blue light radiation
During inspection or replacement of samples, wear protective goggles with
orange lenses.
Cooling of the environmental chamber must always be in operation, when using
a light module.
If cooling is not in operation, the chamber temperature will rise up to approx.
95°C.
NOTE
The light performance of the lamps has its optimum at a working temperature
of 25°C. At higher or lower temperatures inside the chamber, the light current
decreases.
Product Description |
Chapter 5
Model 3900 Series | 5-7