Pin Assignment; Transducer Connection - HBM ME250 S6 Operating Manual

Dual-channel c-f measuring amplifier
Table of Contents

Advertisement

Available languages

Available languages

8
2

Pin assignment

2.1 Transducer connection

S.G. transducers are connected in a four-wire circuit.
Measurement signal (−)
Excitation voltage (−)
Measurement signal (+)
Excitation voltage (+)
Shield
. . . . . . . . . . . . . . . . . . . . .
Fig. 2.1: Pin assignment
The required excitation voltage according to the factory settings is 5 V
(balanced abour earth). You can change it to 1 V with the aid of switch S25.
U
B
5 V
(factory settings)
1 V
(for operation with SD01)
HBM
Pin
Channel 1 Channel 2
. . . . . . . .
30a
11a
. . . . . . . . . .
32a
32c
. . . . . . . .
30c
10a
. . . . . . . . . .
28a
28c
31a+31c
(alternatively 29a+c)
Bridge resistance
R
110 to 4000 
B
R
60 to 4000 
B
RD
BK
WH
BL
YE
Switch S25
ME250S6
A0827-4.1 en/de

Hide quick links:

Advertisement

Chapters

Table of Contents
loading

Table of Contents