Sub-Assemblies, Overview - Siemens MAMMOMAT Novation DR Installation And Startup

Opdima sp
Hide thumbs Also See for MAMMOMAT Novation DR:
Table of Contents

Advertisement

Installation
General
Before starting the installation, a backup of the installation parameters of the
MAMMOMAT Novation
available.
1.
2.
3.
4.

Sub-assemblies, overview

The components shall be placed according to the figure below.
Biopsy
controller
Stand with
X-ray unit
Workstation
Fig. 1 MAMMOMAT 3000 Nova with digital biopsy and spot imaging system Opdimar
NOTE
Siemens AG
Medical Solutions
DR
can be carried out. This may be omitted, if backup already
Connect a PC to the MAMMOMAT Novation
Select Main menu => Backup => Copy installation area to disk => All .
This will store the installation parameters on the hard disk of the PC.
Select Main menu => Backup => Copy installation area to floppy => All .
This will store the installation parameters on the service diskette.
Switch off the MAMMOMAT.
X
X
Y
Y
Z
Z
X
Z
Y
Workstation table
The workstation shall be placed at least 1.5 m from the object
table of the stand.
Biopsy unit
CCD camera
SPB7-230.814.01
Page 1 of 24
Rev. 01
10.04
CS PS 24
DR
and start the service program.