Adjustments / Programs
Collimator, X-Iris
NOTE
WARNING
Prerequisites
•
Attach the center cross in the center of the image intensifier input screen.
•
Connect the service PC to the serial interface on board D1.
•
Start the service program.
Adjustments
•
Select the "Adjustment, Calibration" menu.
•
Answer the query with "yes".
•
Select "X-Iris" in the "Function Groups" combo box.
•
In the window "Available Functions:" select "1. Init (Limits Off / Autosearch)" and click on
Execute.
•
Close the status window by clicking on the OK button.
•
For the query "Do you want to reset the unit," click "No".
•
The X-Iris can now be completely opened and closed with the X-Iris keys.
Siemens AG
Medical Solutions
Because of the mechanical hysteresis of the semi-transparent
slot diaphragm, open the collimator blades to maximum aperture,
prior to adjusting the radiation iris.
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
SPR2-130.061.02
Rev. 11
Page 11 of 30
09.04
CS PS 24
5 - 11
5
5
5
SIREMOBIL Compact
System Manual