Transcutaneous Pacemaker Tests

Output Pulse Measurement
Test Type ............................................................... Quantitative
Measurement Ranges
Output ................................................................. 4.0 mA to 9.99 mA
Accuracy ......................................................... ±2 % of reading or ±50 µA (whichever
Rate .................................................................... 5.0 PPM to 99.9 PPM
Accuracy ......................................................... ±0.5 % of reading or 0.3 PPM
Width................................................................... 1.00 mS to 9.99 mS
Accuracy ......................................................... ±0.5 % of reading or 14 µS (whichever
Energy ................................................................ 1µJ to 999 µJ
Accuracy ......................................................... ±5 % of reading / computation
Demand Mode Test
Test Type ............................................................... Qualitative
Physiological Simulation
Selection ............................................................. Normal Sinus Rhythm (NSR)
Amplitude............................................................ (V
Mode of Operation
Underdrive .......................................................... NSR @ 85 % of measured pulse
Overdrive ............................................................ NSR @ 115 % of measured pulse
Auxiliary Control................................................... The under / overdrive NSR simulations
Introduction and Specifications
Instrument Specifications
10.0 mA to 99.9 mA
100 mA to 250 mA
is greater)
100 PPM to 300 PPM
(whichever is greater)
10.0 mS to 99.9 mS
is greater)
1 mJ to 999 mJ
1.00 J to 1.99 J
generates the complete P-QRS-T
complex.
): 1.0 mV Lead I.
peak
interval / rate
interval / rate
can also be independently adjusted in
1 BPM increments.
1
1-13