Resolution ............................................................. ±1 LSD
Display Format...................................................... 3 digits
Physiological Simulation
Selection (Default) .............................................. Single Channel Simulation (V)
Ventricular Channel ............................................ Simulated R-wave
Width................................................................... 1.0 mS
Amplitude............................................................ 20 mV
Active Outputs .................................................... Ventricular Channel
Available Test Load(s) ........................................ 500 Ω Only
Additional Waveform Selections
Square (SQU)
Triangle (TRI)
Haversine (HSN)
Sine Square (SSQ)
Asymmetrical Triangle (ISO) (Fixed Width2 mS rise time / 13 mS fall time)
Amplitude
Pacer Load Selection.......................................... 500 Ω
Range ............................................................. 0.05 mV
Accuracy ......................................................... ±5 % of setting
Resolution (Step Size).................................... 0.05 mV
Default Setting ................................................ 20.0 mV
Width
Range ................................................................. 0.15 mS to 300 mS
Accuracy ............................................................. 5 % of setting
Selection Count .................................................. 58
Resolution (Step Size) ........................................ 0.05 mS steps from 0.15 mS to
Default Setting (mS) ........................................... 30 mS
Intended Pacemaker Types................................ VVI (Ventricular Pace and Sense
Compatible Pacemaker Rates ............................ 20 to 200 PPM
Introduction and Specifications
Instrument Specifications
Square Wave
peak
(4 mm Banana Jacks) Only
peak
50.0 mV
peak
peak
0.95 mV
peak
0.50 mV
peak
50.0 mV
peak
peak
0.95 mS
1.0 mS steps from 1.0 mS to 19.0 mS
5.0 mS steps from 20 mS to 95.0 mS
25 mS steps from 100 mS to 300 mS
Only)
(50 V
) to
peak
steps from 0.05 mV
peak
steps from 1.0 mV
peak
1
to
to
1-25