Bassam
February 4, 2025
I bought Fluke SMFT 1000 and according to data sheet it measured until 20 A but it couldn't measure more than 14 A and gave me error 14 overload find solve for that
1 comments:
Mr. Anderson
February 10, 2025
The Fluke SMFT-1000 may display an error 14 overload when measuring above 14 A due to the fuse F1 (20 A, 1000 V) being open or failing. The internal test indicates that a failed fuse can prevent accurate current measurement, and replacement must be done by a qualified technician.
Page 2
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
Table of Contents Title Page Introduction........................... Safety Information........................Specifications ..........................Before you Start .......................... Kit Contents ......................... Accessories ......................... How to Use the Rotary Dial..................... Buttons........................... 10 Info Button ..........................11 Display............................ 11 Terminals/Test Leads....................... 12 Error Messages........................13 How to Zero the Test Leads................... 14 Test Setup .............................
Page 4
SMFT-1000 Table of Contents Power AC/DC and Function Tests ................24 Single Phase Inverter Performance Check............24 3-Phase Inverter Performance Check............... 25 AC/DC Voltage Measurement ................26 AC/DC Current Measurement................26 Functional Tests......................27 Insulation Resistance Test (R ).................. 28 Test Method 1 (Keep the Leads)................
Introduction The Fluke SMFT-1000 Multifunction PV Analyzer (the PV Analyzer or the Product) is a battery- operated analyzer for installation testing and periodic inspection of mains-coupled photovoltaic (PV) systems. Table 1 is a list of the main functions. Table 1. Functions...
SMFT-1000 Users Manual Safety Information A Warning identifies hazardous conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test. Specifications Maximum voltage between any Terminal and Earth Ground ..........
Page 7
Multifunction PV Analyzer Specifications Vibration ............. MIL-PRF-28800F: Class 2 Ingress Protection .......... IEC 60529:IP40 Computer Interface........IR (serial) and Bluetooth Wireless connection compatibility..IRR2-BT Accuracy The accuracy specification is defined as ± (% reading + digit counts) at 23 C ± 5 C, ≤80 % RH.
Page 8
SMFT-1000 Users Manual Insulation Resistance R Display Range Measurement Range Resolution Accuracy 0.00 MΩ to 99.99 MΩ 0.20 MΩ to 99.99 MΩ 0.01 MΩ ± (5 % + 5 Digit) 100.0 MΩ to 199.9 MΩ 100.0 MΩ to 199.9 MΩ...
Page 9
± (7.5 % VI + 0.6 V + 0.2 I) Safety SMFT-1000 ............IEC 61010-1 Pollution Degree 2 IEC 61010-2-034 CAT III 1000 V dc, CAT III 700 V ac i100 Current Clamp........IEC 61010-2-032, Type D (for insulated conductors), 1000 V Accessories.............
Page 10
Frequency Range .......... 2.402 GHz to 2.480 GHz Output Power ..........8 dBm SIMPLIFIED EU DECLARATION OF CONFORMITY Hereby, Fluke declares that the radio equipment contained in this Product is in compliance with Directive 2014/53/EU. The full text of the EU declaration is available GlobalTestSupply www.
TL1000-KIT Test Lead Kit IR Optical-to-USB Adapter Cable 6 x AA Alkaline IEC LR6 (for SMFT-1000, not installed) not shown 4 x AA Alkaline IEC LR6 (for IRR2-BT, not installed) 2 x AA Alkaline IEC LR6 (for i100, not installed) GlobalTestSupply www.
Multifunction PV Analyzer Before you Start How to Use the Rotary Dial Use the rotary dial to select the test type. See Table Table 3. Rotary Dial Item Position Description Memory for saved measurements, device settings, and Help MENU information VISUAL Checklist for pre-inspection tests...
SMFT-1000 Users Manual Buttons Use the buttons to control operation of the PV Analyzer, select test results to view, and scroll through selected test results. See Table Table 4. Push Buttons Item Button Description Shows illustrations and instructions for setup and test function ...
Multifunction PV Analyzer Before you Start Info Button The INFO button shows information about how to use each function of the PV Analyzer. As the rotary dial moves to a function, push to see connection drawings and tips about the ...
SMFT-1000 Users Manual Terminals/Test Leads Test leads are connected and stay in place (Keep the Leads) throughout the test. Table 6 shows the input terminals. XW Warning To prevent possible electrical shock, fire, or personal injury, do not use test leads in CAT III or CAT IV environments without the protective cap installed.
Multifunction PV Analyzer Before you Start Error Messages When the Analyzer detects error conditions, the display shows W and an error code. See Table 7. These error conditions disable or stop the test. Tip: Push INFO for instructions about the error message. Table 7.
SMFT-1000 Users Manual How to Zero the Test Leads XW Warning To prevent possible electrical shock, fire, or personal injury, do not use in CAT III or CAT IV environments without the protective cap installed. The protective cap decreases the exposed probe metal to <4 mm. This decreases the possibility of arc flash from short circuits.
Test Setup Use this kit for safety and performance analysis of solar systems according to IEC 62446-1. The kit contains the SMFT-1000 PV Analyzer (PV Analyzer) and the IRR2-BT Irradiance Meter (Irradiance Meter). The PV Analyzer provides safety and performance measurements of the solar system.
SMFT-1000 Users Manual For IV Curve measurements, synchronize the PV Analyzer with the IRR2-BT at the start of the work day: 1. Turn on the PV Analyzer and the Irradiance Meter. 2. Make sure the PV Analyzer and the Irradiance Meter are within the wireless range (<50 m) of each other.
Multifunction PV Analyzer Test Setup Continuity of Protective Earthing and Equipotential Bonding Conductors For accurate measurements, always compensate the resistance of the test leads before you make measurements: 1. Turn on the PV Analyzer. 2. Turn the rotary switch to 3.
SMFT-1000 Users Manual Resistance Test (R The PV Analyzer measures protective conductor resistance ( ) with a test current ≥200 mA (@2 Ω) for: Earthing and Equipotential Bonding Conductors to IEC 62446-1 Clause 6.1 Lightning Protection System (LPS) ...
Multifunction PV Analyzer Test Setup Grounding System To troubleshoot the grounding system with the R Continuous Measurement method: 1. Push for R+ Positive or for R- Negative and follow the on-screen prompts. Polarity Test The polarity test verifies to IEC 62446-1 Clause 6.2 that the positive and negative wires are correctly connected to the solar system combiner box, inverter, or switch gear.
SMFT-1000 Users Manual PV String Open circuit voltage measurement and circuit current test (short circuit test or operational). Open circuit voltage measurement (V Open circuit voltage (V ) measurement to IEC 62446-1 Clause 6.4. This test checks that module strings are correctly wired and that the expected number of modules are connected in series within the string.
Multifunction PV Analyzer Test Setup Voltage/Current Test (V The V is a test to IEC 62446-1 Clause 6.4 for the maximum voltage that the solar panel produces in standard test conditions. The I is a test to IEC 62446-1 Clause 6.5.2 for the maximum current that the solar panel produces in standard test conditions.
SMFT-1000 Users Manual Select PV Model When the Irradiance Meter is not connected, no limits are available and no irradiance or temperature data shows on the display. To take a measurement: 1. Connect the test leads from the PV Analyzer to the solar panel.
Multifunction PV Analyzer Test Setup Quick V Measurement You can take a quick V measurement without connecting the Irradiance Meter or PV model. Pass/Fail limits or Irradiance data do not show with this type of measurement. To take a measurement: 1.
SMFT-1000 Users Manual 4. Push to start the V measurement. The V measurement shows on the display. If the PV model is selected and the Irradiance Meter is connected, the Pass/Fail icons show on the display. The Measure V instructions are grayed-out with a checkmark to indicate that the measurement is done.
Multifunction PV Analyzer Test Setup 8. Push to clear or cancel the dc measurement column and return to the blank state. To make an ac measurement: 1. Connect the test leads to the ac output of the inverter. 2. Connect the Clamp. ...
SMFT-1000 Users Manual 10. Push 11. Push to hold the ac-L3 measurements. The blue column header indicates that the ac-L3 measurements are on hold. The display shows the Efficiency Factor ratio with a Pass or Fail icon. 12. Push ...
Multifunction PV Analyzer Test Setup 3. Connect the clamp to the circuit under test. Tip: Push to see a connection diagram. The PV Analyzer automatically detects if the measurement is ac or dc current. 4. Push to hold the measurement. The measurement is on hold.
SMFT-1000 Users Manual Insulation Resistance Test (R The R Mode is a test for the resistance of the insulation between the ground and the PV Array as required by IEC 62446-1 Clause 6.7. Repeat this test, as a minimum, for each PV array or sub-array.
Multifunction PV Analyzer Test Setup Passed: and a short beep noise indicates the test passed when the results are greater than the preset limits. Failed: and multiple beeps indicates the test failed when the results are less than the preset limits.
You can measure R between any two measurement points in the PV system. This measurement helps to troubleshoot insulation faults on the wiring cables. Fluke recommends that you remove the solar modules for this test as they can influence the result.
Multifunction PV Analyzer Test Setup The dashes show during the measurement calculation and then the test results show on the display: Live Results: measurement results refresh every second. Green check mark appears when result is under the limit. ...
SMFT-1000 Users Manual I-V Curve Test is a test for the maximum voltage that the solar panel can produce under standard test conditions as required by IEC 62446-1 Clause 7.2. I is a test for the maximum current that the solar panels can produce under standard test conditions.
Multifunction PV Analyzer Test Setup Note A warning shows on the display if the PV Analyzer detects reverse polarity at the start of the test. Push to see a connection diagram. When the test is complete, the test results show in the I-V Curve Table: STC column shows values ...
Page 38
SMFT-1000 Users Manual To set limit: a. Use to highlight the options. b. Push to select the highlighted option and edit on new screen. c. Push to save the limit and return to the previous Diode Test.
Multifunction PV Analyzer Test Setup 5. Push to save the results to memory. A confirmation message with ID number shows on the display and then returns to the test screen. Troubleshoot: If the voltage is not within an acceptable range, use the continuous test to find the diode that fails.
SMFT-1000 Users Manual 5. Push to start the measurement. When the measurement is complete, the display shows: measured voltage of blocking diode measured current of blocking diode Passed: and a short beep noise indicate the test passed when the results are greater than the preset limits.
Page 41
If the voltage drop is too low the diode is shorted, if the voltage is “OL” then the diode is open. Tip: Fluke recommends that you repeat the test with reversed polarity (connect yellow test lead to positive anode and connect green test lead to negative cathode. The reading should be always “OL”.
SMFT-1000 Users Manual Surge Protection Device (SPD) Test The SPD test is a check that the device under test (DUT) works as expected. To set: 1. Turn the rotary switch on the PV Analyzer to The display shows the default Bypass Diode test mode.
Multifunction PV Analyzer Test Setup Auto Test Sequence The PV Analyzer has an Auto Test mode that automatically does a test sequence based on these different combinations: With Insulation Test compared to Without Insulation Test Category 1 compared to Category 1 + 2 ...
SMFT-1000 Users Manual 5. Use the to scroll through the tests. The display shows Auto Test Complete message and shows the results. 6. Use to scroll through the results. 7. Push to clear the test results and not save.
Multifunction PV Analyzer Maintenance Figure 2. IR Serial Cable Attachment IR Port IR Port IR Serial Cable IR Serial Cable 3. At the PC, open the TruTest software program. 4. Turn on the PV Analyzer. 5. See the TruTest™ Data Management Software documentation for complete instructions on how to set the date/time stamp and upload data from the PV Analyzer.
SMFT-1000 Users Manual Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. Dirt or moisture in the terminals can affect readings. To clean the terminals: 1. Turn off the PV Analyzer and remove all test leads.
Multifunction PV Analyzer Maintenance Battery Replacement Replace the batteries with six AA batteries. Alkaline batteries are supplied with the Tester. XW Warning To prevent possible electrical shock, fire, or personal injury: Remove the test leads and any input signals before you replace the battery. ...
SMFT-1000 Users Manual Product Disposal Dispose of the Product in a professional and environmentally appropriate manner: Delete personal data on the Product before disposal. Remove batteries that are not integrated into the electrical system before disposal and dispose of batteries separately.
Need help?
Do you have a question about the SMFT-1000 and is the answer not in the manual?
Questions and answers
I bought Fluke SMFT 1000 and according to data sheet it measured until 20 A but it couldn't measure more than 14 A and gave me error 14 overload find solve for that
The Fluke SMFT-1000 may display an error 14 overload when measuring above 14 A due to the fuse F1 (20 A, 1000 V) being open or failing. The internal test indicates that a failed fuse can prevent accurate current measurement, and replacement must be done by a qualified technician.
This answer is automatically generated