Long Term Test - Fluke SigmaPace 1000 Operator's Manual

External pacemaker analyzer
Hide thumbs Also See for SigmaPace 1000:
Table of Contents

Advertisement

SigmaPace 1000
Operators Manual
"S ens ed" C ardiac
T es t P uls e #1
"S ens ed" C ardiac
T es t P uls e #1

Long Term Test

The overall stability of the pacemaker's pulse rate and amplitude are
monitored continuously over an extended time period. Each measured pulse is
compared to factory preset limits of 10 % for both monitored variables. Test
limits are adjustable by the user.
3-8
P ac ing R ate
Interval
R es et > 750 mS
P R P
S R P
P ac ing R ate
Interval
R es et (2 X )
P R P
Figure 3-3. Sensed Refractory Period
"S ens ed" C ardiac
T es t P uls e #2
S R P
eyr004.eps