Demand Mode and Dual-channel (AV) ECG Signal

The following test measures the demand mode pacemaker's ability to interact
with a simulated dual-channel (AV) ECG signal.
1. From the SELECT TEST menu, press F2 DMAND MODE.
2. Press F4 LOAD UP or DOWN arrow key to select the load.
3. Press F1 START to advance the screen and start the test.
4-16