Performance Tests

To prevent possible electrical shock, fire, or personal injury,
do not perform the performance test procedures unless the
Product is fully assembled.
The following performance tests verify the complete operation of the Product and check
the accuracy of each function against the Product's specifications. See Table 2. If the
Product fails any part of the test, calibration adjustment and/or repair is indicated. See
"Calibration Adjustment".
Test
Calibrator Output
(Switch Position)
K
30 V @ 50 HZ
AC Volts
300 V @ 50 HZ
570 V @ 50 HZ
30 V @ 400 HZ
300 V @ 400 HZ
570 V @ 400 HZ
L
0 V
DC Volts
30 V
300 V
570 V
-30 V
-300 V
-570 V
Ω
0
Ω
Ohms
30
300
570
900
3000
5700
XWWarning
Table 2. Performance Tests
29.5 V
297.5 V
565.5 V
29.5 V
297.5 V
565.5 V
-0.3 V
29.6 V
298.2 V
566.9 V
-30.4 V
-301.8 V
-573.1 V
Ω
0
29.6
Ω
298.2
Ω
566.9
Ω
894
Ω
2982
Ω
5669
Detachable Jaw True-rms Clamp Meter
UUT Meter Reading Limit
Low
30.5 V
302.5 V
574.5 V
30.5 V
302.5 V
574.5 V
0.3 V
30.4 V
301.8 V
573.1 V
-29.6 V
-298.2 V
-566.9 V
0.3
Ω
30.4
Ω
301.8
Ω
573.1
Ω
907
Ω
3018
Ω
5731
Performance Tests
High
Ω
Ω
Ω
Ω
Ω
Ω
Ω
7