Fluke SigmaPace 1000 Operator's Manual page 37

External pacemaker analyzer
Hide thumbs Also See for SigmaPace 1000:
Table of Contents

Advertisement

Noise Immunity Test
Test Type ............................................................... Qualitative
Channels
Single.................................................................. Atrial or Ventricular Only
Waveform .............................................................. Sine Wave
Frequency.............................................................. 50 and 60 Hz
Accuracy................................................................ 0.5 Hz
Active Output(s).................................................... Selected atrial and / or ventricular
Output Selections
Atrial Channel Only
Ventricular Channel Only
ECG Signal ............................................................ ECG Signal can be added to the
Amplitude
Pacer Load Selection.......................................... 500 Ω
Range ............................................................. 0.00 (OFF) to 100 mV
Accuracy ......................................................... ±5 % of setting
Resolution....................................................... 5 mV
Pacer Load Selection.......................................... 200 Ω
Range ............................................................. 0.00 (OFF) to 40 mV
Accuracy ......................................................... ±5 % of setting
Resolution....................................................... 5 mV
Pacer Load Selection.......................................... 1000 Ω
Range ............................................................. 0.00 (OFF) to 200 mV
Accuracy ......................................................... ±5 % of setting
Resolution....................................................... 5 mV
Refractory Period Test (Atrial Channel)
Test Selections
Paced Refractory Period (PRP)
Sensed Refractory Period (SRP)
Period..................................................................... 20 to 500 mS
Accuracy................................................................ ±5 % of reading or ±1 mS whichever is
Resolution ............................................................. ±1 LSD
Display Format...................................................... 3 digits
Introduction and Specifications
Instrument Specifications
channel test load
selected Channel
peak-to-peak
peak-to-peak
peak-to-peak
greater
peak-to-peak
steps
peak-to-peak
steps
peak-to-peak
steps
1-23
1