7. Press F3 AMP to adjust the amplitude.
8. Press F4 RATE to adjust the rate.
9. Press F1 START to advance screen and start the test.
Target values for amplitude and rate must be entered for test to
perform correctly.
Transvenous Pacemaker Testing
F1
F2
Note
Setup and Testing
F3
F4
4
eyr051.eps
4-41