To set the reference, do the following:
Turn on the tester and smart remote and let them sit for 1 minute
before setting the reference. Set the reference only after the testers
have reached an ambient temperature between 10
and 104
1
Connect the main and remote testers together as shown in Figure 2.
2
Turn the rotary switch to SPECIAL FUNCTIONS.
3
Highlight Set Reference; then press
4
Press
.
Cat 6A/Class E
RJ45 reference patch cord

Preparing to Save Tests

Check the memory space available:
Turn the rotary switch to SPECIAL FUNCTIONS; then select Memory Status.
Select a cable ID source:
You can select IDs from a pre-generated list or create an ID after each test. Turn the
rotary switch to SETUP, select Instrument Settings, select Cable ID Source; then select a
source.
Set up a job folder:
On the Instrument Settings menu select Current Folder. Select an existing folder or
press
Create Folder to create a new folder.
Enter job information:
On the Instrument Settings menu press
Company names. To enter a new name, select a setting, press
then use the softkeys,
are finished.
F).
channel
A
adapters
6 inch (15 cm)
Figure 2. Twisted Pair Reference Connections
 
Note
.
to show the tab with the Operator, Site, and
, and
C and 40
C (50
Optional
permanent link
adapter
Channel
adapter
Create if necessary;
for editing. Press
F
Y04.EPS