(Example: SETLOAD=0050 is the command for a 50 Ω test load)
Selections for "nnnn" range from 0050 to 1550, in 50 Ω increments.
RDLOAD
Reads the current test load setting. When this command is sent to the Analyzer
it returns: LOAD=nnnn; where nnnn equals the resistance in ohms from 50 to
1550 Ω.
SETTEST
SETTEST=PULSE
The Analyzer monitors the applied pulse rate, amplitude, width, and energy
data.
HOLD
Freezes the pacer measurement and maintains the last reading in memory.
RDMETER
The Analyzer transmits the following information string for rate, amplitude
and width from the data "frozen" from the above HOLD serial command.
The data is returned in the following vertical format: AMP=XXX mA,
RATE=XXX PPM, WIDTH=XXX mS. For each parameter, the Analyzer
returns a three-digit numeric value (XXX) with a floating decimal point to
cover the multiple ranges. The unit of measure remains the same for all
transmitted readings (mA, PPM and mS). Additionally, single parameter
information can be acquired.
RDMETERH
The data is returned in the following horizontal format:
AMP=XXX mA, RATE=XXX PPM.
5-8