Performing Electrical Safety Tests

The Analyzer performs a number of different electrical and
performance tests on biomedical equipment. The following
sections describe the various tests and how to perform
them using the Analyzer.

Setting the Test Standard

The Analyzer performs electrical safety testing based on a
number of different safety standards: AAMI ES1/NFPA99,
IEC 62353, and IEC 60601-1. AAMI is the Analyzer's
default standard.
To select another standard:
1. Press the softkey labeled Standard to open the scroll
box above the softkey label.
2. Press  or  to scroll through the standard
selections.
Some electrical tests may not be applicable for a specific
standard. In these cases, the Analyzer's menu will not
display the excluded test as a selection.
12

Performing Mains Voltage Testing

The Mains Voltage test measures the voltage on the mains
input through three separate measurements. To access
the Mains Voltage test, press . A Mains Voltage test is
shown in Figure 6.
Figure 6. Mains Voltage Test
fis204.bmp