MRU GmbH
Plug the viton tubes on the filter unit. Push the filter unit on the clip.
Position 2 and 3: Dust- and particle filter (#65533 and #66088)
Required materials: Dust- and particle filter (#65533 and #66088) contained in the
service set.
Required tools: needle-nose pliers.
Exchange interval: Exchange necessary, if the filter gets dark or black.
Steps:
Pull the viton-tubes from the filter unit with the hand. If it is necessary, pliers can be
used to solve the tubes from the filter unit.
Remove the
exhausted filter
unit.
Remove the exhausted filter unit from the clip.
Plug the viton tubes on the filter unit. Push the filter unit on the clip.
Install a new filter
unit.
SWG 100 BIO Ex User Manual
Page 119 of 150
Need help?
Do you have a question about the SWG100 BIO-EX and is the answer not in the manual?