Instructions For Use - Otto Bock 17B66 Series Instructions For Use Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 9
NOTICE
Thermal overloading of the product
Damage due to improper thermal treatment
► Do not carry out any heat treatment at temperatures above 300 °C
(570 °F).
► Prior to thermal treatment, remove all temperature-critical components
(such as plastic parts).
NOTICE
Exposure of the product to unsuitable environmental conditions
Damage, brittleness or destruction due to improper handling
► Avoid storage in condensing ambient humidity.
► Avoid contact with abrasive substances (e.g. sand, dust).
► Do not expose the product to temperatures below -10 °C (14 °F) or
above +60 °C (140 °F) (e.g. sauna, excessive sunlight, drying on a
radiator).
5 Preparing the product for use
INFORMATION
Parallel alignment of the orthotic joints
Use the 743R6 alignment fixture to position the orthotic joints on the plaster
positive.
Contouring the bars and foot stirrups
Various foot stirrups are available for the orthotic ankle joint.
Process the bars and foot stirrups according to their instructions for use.
Gluing in bars
Required materials: 636W28* special adhesive, degreasing cleaner
The bars must be glued in the insertion zones to ensure stability.
1) Clean the connecting surfaces with a degreasing cleaner.
2) Apply the special adhesive to the connecting surfaces.
3) Insert the bars.
4) Secure with the enclosed screws.
5) Allow to cure for at least 4 hours.
INFORMATION: Final bonding strength will be reached after
16 hours.