5.3.2

Absorbance

5.3.3

Glow Type Luminescence

June 2002
Operating Manual for GENios, GENios FL, and GENios Plus No: I 112 904 Rev No: 1.1
The specifications only apply to 96 well plates.
Absorbance can not be measured in 1536 well plates.
The absorbance filters are not within the instrument's standard configuration.
Parameters
Characteristics
Wavelength
230 - 1000 nm
range
Meas. range
0 - 3 OD
Resolution
0.0001 OD
Accuracy
0.0 - 2.0 OD
2.0 – 3.0 OD
Precision
0 - 3.0 OD
Linearity
0 - 3.0 OD
No. of Flashes /
1 - 255
Well
Standard filters
405 nm / 450 nm / 492 nm / 620 nm
(not included in standard delivery)
The specifications only apply to 96 well plates.
Luminescence can only be measured in 96 well plates.
Parameters
Sensitivity
Wavelength range
Gain
Integration setting
Manual
Maximum Dynamic
Range
±
1 % and
±
1.5 % and
±
1 % and
±
1.5 % and
Characteristics
-6
< 0.4 10
DEA activity units/well of molecular biology
grade alkaline phosphatase (20 µl/well) using
CDP-Star (Tropix) chemiluminescent substrate mix
with Emerald-II enhancer (100 µl/well) in a Greiner
96 well white plate. Use of the cover plate greatly
reduces the cross-talk.
-6
0.4 10
DEA activity units/well correspond to 200 fg/
well of alkaline phosphatase when using a lot of
2000 DEA activity units/mg protein.
400 - 700 nm
Optimal:
preset value
Manual:
1 - 255
Integration time
1 - 5000 ms
Automatic
(1 - 1000 ms)
5. Instrument Features
±
0.010 OD
492 nm
±
0.010 OD
492 nm
±
0.005 OD
492 nm
±
0.005 OD
492 nm
Measurement range
0 - 55,000 RLU
0 - 55,000,000 RLU
(using XF
)
LUOR
5-11