Transformer Test - Applied Precision WS 2120 User Manual

Table of Contents

Advertisement

Meter Readout page enables to read content of registers of multifunctional meters with optical
communication interface according to IEC 62056-21.
Optical Communication head OPTH 1200 has to be connected to AUX input of the device and properly
placed to meter under test before starting data readout.
Meter Readout is initiated by F4 key and result (content of registers of meter) is printed on display.
Readout data can be saved into database using key F4.
Meter Readout screen
Test of current transformers (CT) or voltage transformers (VT) can be done after connecting of suitable
combination of current and voltage measuring sensors.
Testing of single phase transformer can be done using one three phase measuring accessory which is
connected to transformer this way: L1 channel of accessory to secondary side and L2 (or L3) channel
to primary side. Measurement screen shows result for both variants (L2/L1 and L3/L1).
Testing of three phase transformer (or also single phase transformer) can be done using two three
phase current sensors (in case of current transformer testing) resp. two three phase voltage sensors
(in case of voltage transformer testing). Channels of sensor connected into input A has to be connected
to secondary side of transformer and channels of sensor connected into input B has to be connected
to primary side of transformer.
It is possible to use any combination of current measuring accessories (CT 2x20, CC 2x12, CC 2x24,
FCP 3xxx) for Current Transformer testing.
Measurement
screens
nominal and measured Transmission Ratio, Transmission Error and measured Phase Error.
www.appliedp.com
4.9 Meter Readout
Content:
(data) .............. Data from meter or error message
Control keys:
ESC ..................... Exit (return to previous page)
F3 .......... Read data from meter
F4 .......... Save data into database
F5 .......... Menu
▼▲ ....... Scrolling of readout data down and up
◄ ► ...... Scrolling of readout data left and right
PgUp, PgDn ... Paging of readout data up and down
* P48

4.10 Transformer Test

depend
on
combination
WS-UGFW619-EN
Working Standard
of
connected
sensors
User's Guide
and
show:
45

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ws 2320

Table of Contents