Applied Precision WS 2120 User Manual page 82

Table of Contents

Advertisement

Working Standard
User's Guide
SYST:ENER:DOS:MAXD:SUBP {<subperiod>}
Setting the value of Maximum Demand Test's sub-period parameter.
SYST:ENER:DOS:DUT:MAXD:K {<register>}
Setting the register value of Maximum Demand Test for DUT (Device Under Test).
SYST:ENER:DOS:RES:MAXD?
Request for error of Maximum Demand Test after finishing this test.
SYST:ENER:DOS:STAT:MAXD:K?
Request for state of Maximum Demand Test.
SYST:ENER:DOS:STAT:MAXD:K:TEMP?
Request for the recently integrated value of power (Maximum Demand Test).
SYST:ENER:DOS:STAT:MAXD:WIN:BEG?
Request for the beginning timestamp of Maximum Demand Test being in progress.
SYST:ENER:DOS:STAT:MAXD:WIN:END?
Request for the ending timestamp of Maximum Demand Test being in progress.
SYST:DOS?
Request for state of Dosage mode. Answer is 0 (=disabled) or 1 (=enabled).
SYST:DOS:ENAB {enable(0|1) }
Setting of state of Dosage mode. Parameter enable can be 0 (=disable) or 1 (=enable).
SYST:DOS:FALL?
Request for Dosage fall method. Answer is 0 (=soft generate) or 1 (=immediate).
SYST:DOS:FALL { method(0|1) }
Setting of Dosage fall method. Parameter can be 0 (=soft generate) or 1 (=immediate).
SYST:DOS:RISe?
Request for Dosage rise method. Answer is 0 (=soft generate) or 1 (=immediate).
SYST:DOS:RIS { method(0|1) }
Setting of Dosage rise method. Parameter can be 0 (=soft generate) or 1 (=immediate).
SYST:DOS:STOP:VOLT { state (0|1) }
Setting of voltage Stopping feature after Dosage completion. Parameter can be 0 (disabled) or 1
(enabled).
SYST:DOS:STOP:VOLT ?
Request for state of "voltage Stopping feature after Dosage completion". Answer can be 0 (disabled) or
1 (enabled).
SYST:DOS:STOP:CURR { state (0|1) }
Setting of current Stopping feature after Dosage completion. Parameter can be 0 (disabled) or 1
(enabled).
SYST:DOS:STOP:CURR ?
Request for state of "current Stopping feature after Dosage completion". Answer can be 0 (disabled) or
1 (enabled).
82
www.appliedp.com
WS-UGFW619-EN

Advertisement

Table of Contents
loading