Meter Error Test Parameters shown in Table of Parameters of selected test:
DATE
TIME
METERERROR
METERSTDEV
METERCONST
METERUNIT
OPSINPUTTYPE
OPSINTTIME
TESTSAMPLES
MDFIXActive;
MDFIXApparent
MDSlidActive;
MDSlidApparent
MDPeriod
MDSubPeriods
ENERGYTYPE
CONTYPE
MSINPH
MACTCURR
MREACTCURR
MZ
MPHZ
www.appliedp.com
test date
test time
test result
standard deviation of result
meter constant value
meter constant unit
input type
integration time [s]
number of samples
MD power register actualized
in fixed time window
MD power register actualized
in "slided" time window
Size of time window
(Fix/Slide) of MD test
Number of sub periods of MD
Test
energy type
connection type
Sinus of phase shift between
current and voltage (3x)
Ohmic load of current
component (3x)
Reactive component of
current (3x)
Load impedance (3x)
Phase shift of impedance
load (3x)
FREQ
MCURR
MVOLT
MPHI
MPHU
MDFI
MDFU
MPOWER
MQPOWER
MSPOWER
MPOWFACT
MENERGY
MQENERGY
MSENERGY
MSPOWERSUM
MPOWERSUM
MQPOWERSUM
MSENERGYSUM
MENERGYSUM
MQENERGYSUM
MVOLT12
MVOLT23
MVOLT31
MPHU12
MPHU23
MPHU31
WS-UGFW619-EN
Working Standard
User's Guide
frequency
current (3x)
voltage (3x)
current phase angle (3x)
voltage phase angle (3x)
current distortion (3x)
voltage distortion (3x)
active power (3x)
reactive power (3x)
apparent power (3x)
power factor (3x)
active energy (3x)
reactive energy (3x)
apparent energy (3x)
sum of apparent powers
sum of active powers
sum of reactive powers
sum of apparent energies
sum of active energies
sum of reactive energies
line voltage between L1;L2
line voltage between L2;L3
line voltage between L3;L1
phase shift between line
voltage L12 and L1
phase shift between line
voltage L23 and L2
phase shift between line
voltage L31 and L3