Applied Precision WS 2120 User Manual page 92

Table of Contents

Advertisement

Working Standard
User's Guide
Request for measured samples of the waveform. Returns first 64 values from specific channel (phase
(0=L1, 1=L2, 2=L3), unit (0=U, 1=I)). Answer: amplitude samples (64 comma separated real
numbers).
MEAS:SIGN:AMPL? {<phase (0|1|2)>},{< unit (0|1)>}
Request for measured amplitude values of higher harmonic content. Returns first 40 values from
specific channel (phase (0=L1, 1=L2, 2=L3), unit (0=U, 1=I)). Answer: 40 comma separated real
numbers.
MEAS:SIGN:PHAS? {<phase (0|1|2)>},{< unit (0|1)>}
Request for measured phase shifts of higher harmonic content. Returns first 40 values from specific
channel (phase (0=L1, 1=L2, 2=L3), unit (0=U, 1=I). Answer: phase shifts (40 comma separated
numbers).
MEAS:OPER {<param>}
Setting of Operation Mode. Parameter is Operation Mode's number (0-7) or code (P6-K3).
MEAS:OPER?
Request for setting of Operation Mode. Replay is Operation Mode's number (0-7).
MEAS:CONN:TYP {<param>}
Connection setting. Parameters: 0 - 3P4W, 1 - 3P3W, 2 - 2P3W, 3 - 1P3W, 4 - 1P2W
MEAS:CONN:TYP?
Request for connection setting. Answers: 0 - 3P4W, 1 - 3P3W, 2 - 2P3W, 3 - 1P3W, 4 - 1P2W
MEAS:CHNLA:AC?
Request for values from Input A.
MEAS:CHNLA:AC:PHAS?
Request for values of phases from Input A.
MEAS:CHNLA:AC:SUM:SQ?
Request for sum of square voltages or currents from Input A.
MEAS:CHNLA:AC:DIST?
Request for distortion of measured voltage or current from Input A.
MEAS:CHNLA:MOD?
Request for accessories type connected to Input A.
Answer can be: 2 ... voltage, 1 ... current, 0 ... none.
MEAS:CHNLB:AC?
Request for values from Input B.
MEAS:CHNLB:AC:PHAS?
Request for values of phases from Input B.
MEAS:CHNLB:AC:SUM:SQ?
Request for sum of square voltages or currents from Input B.
MEAS:CHNLB:AC:DIST?
Request for the distortion of measured voltage or current from Input B.
92
www.appliedp.com
WS-UGFW619-EN