Applied Precision WS 2120 User Manual page 81

Table of Contents

Advertisement

Working Standard
User's Guide
SYST:ENER:DOS:TYP {<type>}
Setting of Dosage Type. Parameter is 0 for time or 1 for energy.
SYST:ENER:DOS:TYP?
Request for state of Dosage Type. Answer is 0 for time or 1 for energy.
SYST:ENER:DOS:TIM {<time>}
Setting of time (in seconds) for time dosage.
SYST:ENER:DOS:TIM?
Request for time for "time dosage". Answer is time in seconds.
SYST:ENER:DOS:ENER {<energy>}
Setting of energy (in Ws or VAs) for energy dosage.
SYST:ENER:DOS:ENER?
Request for energy (in Ws or VAs) for energy dosage.
SYST:ENER:DOS:ENER:K {<energy>}
Setting of energy (in kWh or kVAh) for energy dosage.
SYST:ENER:DOS:ENER:K?
Request for energy (in kWh or kVAh) for energy dosage.
SYST:ENER:DOS:ENER:TYP {< type >}
Setting of energy type for Energy Dosage (Test).
Parameters are 0 = Active, 1 = Reactive, 2 = Apparent.
SYST:ENER:DOS:ENER:TYP?
Request for energy type for Energy Dosage (Test).
Answers are 0 = Active, 1 = Reactive, 2 = Apparent.
SYST:ENER:DOS:RES:SAV
Saving the results of Dial (Dosage) test. Return values are: 0 – result available, 1 – saved result, 2 –
no result available.
SYST:ENER:DOS:RES?
Request for overall error of Dial (Dosage) test after finishing this test.
SYST:ENER:DOS:DUT:ENER:K {<x>,<y>}
Setting the values of Dial Test from el. meter. Parameters are: x = value type, 0 – Initial, 1 – Final, 2 –
their difference. y = actual value of chosen type by x.
SYST:ENER:DOS:DUT:ENER:K?
Request for all values of meter dial: Initial, Final, and differential (delta).
SYST:ENER:DOS:MAXD:WIN {<window>}
Setting the value of Maximum Demand Test's window parameter. This parameters represents setting
of Sliding or Fixed type of window.
SYST:ENER:DOS:MAXD:PER {<period>}
Setting the value of Maximum Demand Test's period parameter. This parameters represents time
stated in minutes.
www.appliedp.com
81
WS-UGFW619-EN

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ws 2320

Table of Contents