Vector Diagram; Meter Error Test And Meter Constant Test - Applied Precision RS 2130 User Manual

Reference standard, standard meter of electrical power and energy
Table of Contents

Advertisement

Vector diagram can be activated by key DEL. In this diagram are showed all voltages and currents
(minimal amplitudes are 0.5V, 0.5mA).
Diagram represents graphic interpretation of harmonic content. There is a chart on the left and by up
and down arrows zooming is changed in range 1x - 20x. On the right side there is a list of harmonic
st
items where the 1
column from left (Harm) means the number of harmonic item, 2
states for harmonic item's amplitude in % related to the fundamental harmonic and the 3
states for phase shift of harmonic item.
Paging of this list is possible via left and right arrows on the keypad. Repeated pressing of DEL button
rotates the currently displayed channel. There are 6 available channels:
L1 VOLTAGE, L1 CURRENT ... L3 CURRENT.

5.15 Meter Error Test and Meter Constant Test

Meter error can be measured in OPS Measurement Mode. This mode is used to control Local
Evaluation Unit OPS and measure error of connected electricity meter. This mode can be selected
using menu item Measurement / Energy / Impulse / OPS Meas/Test -> or by pressing "EXP" key from
the basic mode if device has installed appropriate hardware (hardware versions 1.08 and more). Mode
can be closed by pressing "ESC" or "EXP" buttons.
The basic activities of the device are not suspended – device is measuring in background.
Description of OPS Measurement Mode window:
st
rd
1
- 3
Line:
- measured values .... measured values of (active, reactive or apparent) power, voltage and
th
th
4
- 5
Line:
- mode indicator ....... mode indicator shown in the beginning of 5
- impulse counter ..... counter of impulses captured from tested meter shown in the beginning of
- sample counter ...... counter of samples shown in the beginning of 5
- error / state ............. measured meter error or actual state (mode) of local evaluation unit OPS
www.appliedp.com

5.14 Vector Diagram

P: 95.37 , 95.42 , 95.21
U: 230.3 , 230.6 , 229.8
I: 414.1m, 413.8m, 414.3m |<3>Input LED
3
-0.017%
e:06/10
k:10000
▲02:-0.012% |<6>Imp.
AVG:-0.013%
03:-0.015% |<7>Samp. 10
STD:+1.23E-02 ▼04:-0.014% |<8>Mode<9>Clr
OPS Measurement Mode window – Meter Error Test
Left part of window
current in all channels.
test and "k" indicates meter constant test.
th
4
line. This number has initial value equal to value of right part parameter
"Imp." and is lowered by 1 after every received impulse. When it reaches
zero then new measured sample is calculated.
Counter shows progress: number of measured samples / required number
of samples set in right part parameter "Samp.".
is shown with large letters in both lines
Reference Standard RS 2x30, RS 1x30
|<1>Meas
|<2>Calib OFF
|<4>Res.
|<5>Wind. 10%
RS-UG91-EN
User's Guide
nd
column (U[%])
column (ph[°])
rd
ON
HIGH
5
line: "e" indicates meter error
th
th
line after mode indicator.
35