Applied Precision WS 2120 User Manual page 43

Table of Contents

Advertisement

Add meter screen from Test Result
Details screen
Optical Sensor Calibration has to be done in case of disk mark of poor quality, week emitting LED of
meter or stronger environmental influence. Calibration process is described in Accessories' User Guide
in chapter Optical Sensor.
Calibration is activated by key F2 from Meter Error Test page.
Control keys:
F2 ........... Exit (return to previous page)
ESC ........ Exit (end of Meter Error Test)
www.appliedp.com
Content:
SN ....................... Serial number of meter
Type .................... meter type
Connection ......... connection type regarding wires
Input Type .......... error registration method
Energy ................ energy type during test
Constant ............ value of the constant for the selected
Constant unit ..... unit of constant
Constant Type ... primary or secondary
Producer
Class ................... accuracy class
Year ..................... of production
Control keys:
* P90
4.7.2 Optical Sensor Calibration
WS-UGFW619-EN
number
registration method
F1, ESC .. Exit (return to previous page)
F4............. Saves all parameters to newly defined
meter
Working Standard
User's Guide
43