Little back triangle in top right corner in one of the voltage or current values determines the
synchronization frequency channel.
Harmonics View screen shows measured values (amplitude and phase) of all harmonics.
Harmonics View screen
Control keys (for three graphical presentation pages):
F1 ...... Switch to another graphical presentation pages or back to main page
F2 .......................... Switch to Meter Error Test
F3 .................. Show Harmonics View in Harmonics screen, switch to Transformer Test
F4
...................... Save Load Values
F5
...................... Menu
▼▲ ....................... Choose combination of showed signals in Harmonics and Curve screens
◄ ► ...................... Scroll next/previous 10 harmonics in Harmonics screen
ESC ....................... Exit (return to main measurement page)
Control keys (for Harmonics View screen):
F1, ESC ............Exit (return back to Harmonics screen)
F2 .....................Switch to next channel (L1
F3 .....................Switch to next channel (U
F4 .....................Save Harmonics
▼▲◄ ► ...........Scrolling of content
.......Paging of content
PgDn, PgUp
26
Individual harmonic components are distributed over several
screens (3 x I, 3 x U). Each screen can be scrolled up and down.
Individual amplitudes represent a percentage of the first
harmonic and phase shift
corresponding harmonic component. At the first harmonic the
phase shift is given in brackets (measured against first harmonic
in first voltage phase) but otherwise it is 0, which serves as a
reference for determining the phase difference of other
harmonics.
* P65
<->
WS-UGFW619-EN
is relative displacement
L2
L3
L1)
<->
<->
<->
I
P
U)
<->
<->
Working Standard
User's Guide