Measured samples can be seen in Test Result Details page which can be activated by F3 from Meter
Error Test page. This screen shows all test parameters and all measured samples with pulses or
energies against time used for evaluation of these samples.
Samples screen
Up to version 5.30 (including) of FW the quantity E[xxx] / Time [s] with different meaning „calculated energy over time
1
*
with respect to Energy and Error [%]".
Control keys:
F1, ESC .. Exit (return to previous page)
F3 ............. Enables to store recent test results to new meter (Add meter screen will appear)
F4 ............. Saves recent test results to database for currently selected meter
F5 ............. Toggles between Exp. / Meas imp. and E[kWh] / Time [s] view in third column
42

4.7.1 Test Result Details

Content:
SN .................... serial number of meter
Error ................. resulting error
Deviation ......... standard deviation of last completed test
Impulses .......... number of (required) impulses
Energy ............. energy being measured during last
Samples:
No ..................... Sequence number
Error [%] ........... Measured sample (error in %)
The last column toggles between (F5):
Exp. / Meas imp. ...... Expected impulses count / Measured
E[kWh] / Time [s] ..... measured energy over time (unit
* P66
WS-UGFW619-EN
Working Standard
completed test
impulses count or
or
changes with Measurement Settings)
User's Guide
*1
www.appliedp.com